Calibration of High Dose Rate Cobalt-60 Source Using a Farmer Type Chamber in A Locally Constructed Hollow Phantom

Subscribe to access this work and thousands more

ABSTRACT

An important aspect of general quality assurance program for brachytherapy dosimetry is source calibration. Vendors assign large uncertainties of up to ±10 % for the stated source calibrated values. It is thus important to check the vendor stated calibration. There are three main methods of calibrating brachytherapy sources and they are the “Free- in – air” method, using a re-entrant(well ) type chamber and using a thimble (Farmer) type ionization chamber in a phantom. The main objective of this study was to compare values of source strength obtained by using the re-entrant chamber method and the thimble (farmer) type ionization chamber in a phantom method. A cylindrical hollow phantom was constructed from Poly methyl methacrylate (PMMA) having four holes located at a distance of 8cm from the centre and the angle between each of the holes being 90° from the centrally located source holder . The experimental set up followed International Atomic Energy Agency (IAEA) Technical Document 1274 (IAEA-TECDOC-1274) for the re-entrant chamber method and DIN 68909-2 in combination with DGMP report 13 for the thimble ionization chamber in a phantom method. Measurements carried out using the well type chamber deviated from the theoretical value of the source by -1.36% whereas those measured by the Farmer type ionization chamber in a locally constructed phantom deviated from the theoretical air kerma strength value by -2.1%. The difference in measured values from the two measuring methods was found to be0.8%.